TS120冷熱沖擊試驗(yàn)箱

價(jià)格
電議

型號(hào)
TS120

品牌
德國WEISS

所在地
暫無

更新時(shí)間
2019-02-08 08:05:03

瀏覽次數(shù)

     


    Temperature range of
    hot chamber

     

    from +50 °C to +220 °C
     
    Temperature range of
    cold chamber 
    from -80 °C to +70 °C
     
    Hot or cold chamber may also be utilized as individual temperature chamber.
     
    Temperature deviation in time
    in centre of working space

    Temperature deviation in space 
    ±0.3 K to ±1.0 K

    ±0.5 K to ±2.0 K
     
    Heating rate of hot chamber according to IEC 60068-3-5
     
    14 K/min 
    Cooling rate of cold chamber
    according to IEC 60068-3-5
     
    63 K/min.
     
    Heating rate of cold chamber
    according to IEC 60068-3-5
     
    2 K/min.
     
    Temperature changing tests:
    MIL 883 E Method 1010.7
    Severity D with 12 kg of IC&s
    distributed over 3 shelves;
    measurement on test speci-
    mens

     

     


    Adaption time < 15 min.
     
    The following standards are met
    DIN 40 046 page 14 test Na
    transfer time < 10 sec.;
    IEC-60068-2-14 test Na
    transfer time < 10 sec.;
    MIL-STD-810 E method 503;
    MIL-STD-883 E method 1010.7
    variants ABCDF
    JEDEC standard JESD22-A101-A
     
    Temperature calibration values are measured at 
    - 40 °C cold chamber
    + 125 °C hot chamber 

     

    Pre-tempering Temperature change rates can be
    increased by choice of raised or
    lowered temperatures in warm/cold
    chamber. 

     

     

    Data for installation and operation ...
     
    Test space dimensions   470 mm wide
      650 mm deep
      410 mm high
     
    Dimensions of installation and operation 
      970 mm wide (without colour-
                                    touchpanel)
    1290 mm wide (with colour-touchpanel)
    2350 mm deep
    2450 mm high (with cover for
                                    ball spindle)
     
    Min. possible dimensions for transport (Additional expenses possibly caused by disassembly of corresponding parts)   970 mm wide
    2190 mm deep (doors demounted)
    1985 mm high
     
    Operating conditions Ambient temperature +10 °C
    to +35 °C; max. rel. air humidity
    75 % r. h. max. dew point +20 °C
     
    Installation conditions The max. permissible ambient temperature for storage
    is -25 °C to +55 °C.
    Note:
    There must not be any residual water like e.g. humidification or cooling water in the cabinet!

    Please protect test chamber against direct sunlight and
    sources of heat.

     

     

    以上信息由企業(yè)自行提供,信息內(nèi)容的真實(shí)性、準(zhǔn)確性和合法性由相關(guān)企業(yè)負(fù)責(zé),儀器儀表交易網(wǎng)對(duì)此不承擔(dān)任何保證責(zé)任。
    溫馨提示:為規(guī)避購買風(fēng)險(xiǎn),建議您在購買產(chǎn)品前務(wù)必確認(rèn)供應(yīng)商資質(zhì)及產(chǎn)品質(zhì)量。

    其他推薦產(chǎn)品

    WKS3-480/70/10快速溫變?cè)囼?yàn)箱
    WKS3-480/70/
    電議
    WKS3-340/70/5快速溫變?cè)囼?yàn)箱
    WKS3-340/70/
    電議
    WKS3-270/70/25快速溫變?cè)囼?yàn)箱
    WKS3-270/70/
    電議
    WKS3-270/70/20快速溫變?cè)囼?yàn)箱
    WKS3-270/70/
    電議
    WKS3-270/70/15快速溫變?cè)囼?yàn)箱
    WKS3-270/70/
    電議
    WKS3-270/70/10快速溫變?cè)囼?yàn)箱
    WKS3-270/70/
    電議
    WKS3-270/40/10快速溫變?cè)囼?yàn)箱
    WKS3-270/40/
    電議
    您是不是在找:
    恒溫恒濕試驗(yàn)箱鹽霧試驗(yàn)箱高低溫試驗(yàn)箱濕熱試驗(yàn)箱老化試驗(yàn)箱砂塵試驗(yàn)箱步入式試驗(yàn)箱步入式實(shí)驗(yàn)室氙燈耐氣候試驗(yàn)箱箱式淋雨試驗(yàn)箱冷熱沖擊試驗(yàn)箱高低溫交變濕熱試驗(yàn)箱其它試驗(yàn)箱

    首頁| 關(guān)于我們| 聯(lián)系我們| 友情鏈接| 廣告服務(wù)| 會(huì)員服務(wù)| 付款方式| 意見反饋| 法律聲明| 服務(wù)條款

    国产自产v一区二区三区c按摩_亚洲AV秘片一区二区三3_色婷婷久久综合中文久久一本_色婷婷综合久久久久中文字幕